Meet Inspiring Speakers and Experts at our 3000+ Global Conference Series Events with over 1000+ Conferences, 1000+ Symposiums
and 1000+ Workshops on Medical, Pharma, Engineering, Science, Technology and Business.

Explore and learn more about Conference Series : World's leading Event Organizer

Back

Allibardi S

Copan Italia, Italy

Title: Performance of a New Home-Based Self-Vaginal Device for Diagnosis of Chlamydia trachomatis and Neisseria gonorrhoeae

Biography

Biography: Allibardi S

Abstract

Background: self vaginal sampling is a new collection approach for detection of Sexually Transmitted Infections and is able to guarantee privacy and comfort during the collection. The aim of the study was to evaluate usability, vaginal cells collection efficiency and ability to preserve nucleic acids stability of a new self vaginal flocked swab (FLOQSwabTM, Copan) developed for home collection.

Methodology: 80 donors (18 to 45) performed a double self-vaginal sampling (n=160) using: a certified flocked self-vaginal point of care collection (POC) device as a reference method (Copan); a new home-based self-vaginal flocked swab by following the kit instructions. Patients received a questionnaire to assess the usability of the new device. Home-based and POC swabs were processed using Xpert CT/NG® assay (Cepheid). The threshold cycle value (Ct) of a human genomic target, Ct of pathogens (Chlamydia trachomatis CT and Neisseria gonorrhoeae GC2-GC4) and extraction and amplification control (Bacillus globigii spores) were considered to compare performance between the two devices. To evaluate the stability of the nucleic acids at time 0 and after 4 weeks of storage at + 4 ° C and + 30 ° C, 54 negative home collected samples were inoculated with a suspension of CT and GC ATCC (VR880-43069) at 1 and 10xLOD of molecular assay. Findings: 100% of overall agreement was obtained comparing the two devices: 77/80 negative and 3/80 CT positive patients were detected. No failure results were observed. The survey reported a better appreciated home-based collection (80%) with respect to the POC sampling. After 4 weeks of storage at 4°C and at 30°C all spiked samples were detected. Conclusions: the new home-based device has shown the same performance of the reference swab, demonstrating an efficient recovery of vaginal cells, stability of CT and GC nucleic acids up to 4 weeks and excellent acceptability by women